Radionuclide Guide

Tb-161 · Terbium-161

Beta emitter with abundant conversion and Auger electrons. Half-life 6.953 days.

Tb-161 behaves much like Lu-177: a similar half-life, the same chelating chemistry, and beta emission. It also emits many low-energy conversion and Auger electrons, which deposit their energy over very short distances, and its photons are lower in energy. Day-to-day handling looks familiar; measurement settings and surveys need attention.

What changes in practice

  • Handled much like Lu-177, with the same chemistry and a similar half-life.
  • Photons are low in energy (about 48 and 75 keV), so meters with a poor low-energy response under-read. Check the energy response.
  • The dose calibrator needs a setting for Tb-161, and low-energy photons make the reading sensitive to the container. Use a factor for the actual vial and syringe.
  • Produced from enriched Gd-160 without carrier, so there is no long-lived metastable impurity like the Lu-177m found in some Lu-177.

Decay chain

  1. Tb-161 · 6.953 d

    Beta minus to stable Dy-161

Main emissions

TypeFromEnergyPer Tb-161 decay
ElectronsTb-161197 keV mean electron energy per decay (beta, conversion and Auger)every decay
ElectronsTb-161conversion and Auger electrons above 3 keVabout 2.27 per decay
GammaTb-161about 48 keV (several lines)43.1%
GammaTb-16174.6 keV10.3%

Detection and measurement

  • Contamination: a GM pancake detects the betas well.
  • Low-energy photon probes (thin NaI) also work; check what the meter reads at 48 and 75 keV before relying on it.
  • Dose calibrator: establish the Tb-161 setting with the supplier’s reference and your own containers.
  • Imaging: SPECT on the 48 and 75 keV photons is possible, with energy windows set for them.

Shielding

  • Low-energy photons are stopped well by thin lead; standard syringe and vial shields designed for Lu-177 are expected to be adequate, but confirm with a survey.
  • Use plastic (acrylic) close to beta sources, with lead outside it if needed.

Contamination control

  • As for Lu-177: gloves, absorbent liners, and hand and clothing surveys before leaving the area.
  • Conversion and Auger electrons matter for dose to cells, not for external shielding. Skin contamination should still be removed promptly.

Waste and decay-in-storage

  • Ten half-lives is about 70 days, within the NRC 120-day limit for decay-in-storage (10 CFR 35.92). Agreement States may differ.
  • Check the supplier certificate for radionuclidic impurities before planning release.

Licensing

  • The license must authorize Tb-161 in the form used. A license amendment and Authorized User review may be needed before the first patient.

Compared with Lu-177

  • Half-life 6.95 d against 6.65 d: scheduling, release planning and decay-in-storage look much the same.
  • Same chelators and targeting molecules, so radiopharmacy steps are similar.
  • More conversion and Auger electrons per decay; the radiobiology differs, not the shielding.
  • Lower-energy photons: a different dose calibrator setting, and meters that under-read at low energies.

Sources

Decay data checked against these sources in October 2026. Practice notes are general guidance for planning; your license, your procedures and your Radiation Safety Officer decide how Tb-161 is handled at your facility.